B6-C
Product Description
Purpose
Thickness measurement of products from metals and their alloys, thickness measurement under protective coatings, thickness measurement of non-metallic products (plastic, ceramics and other), which allow use of ultrasonic inspection methods.
Features
— colour display (TFT 2.4” 320 × 240);
— possibility of graphic display of ECHO signal in A-scan mode;
— B-scan;
— possibility of metal thickness measurement under coating (in ECHO-ECHO mode and PULSE-ECHO-COATING mode);
— possibility of coating thickness measurement (in PULSE-ECHO-COATING mode);
— automatic probe recognition.
Basic technical characteristics
— range of measured thicknesses (for steel)* Т, mm: with dual transducers: 0.5–300; with direct transducer (with delay line): 0.25–10 for steel; with specialized dual transducers (for measurements under protective coatings with thickness** up to 2.0 mm): 0.25–100;
— speed of ultrasound propagation, m/s: 1000–9999;
— measurement resolution, mm: 0.1; 0.01;
— measurement accuracy, mm: ±(0,005T + 0,05) in range of 0,25 to 10,0; ±(0,01T + 0,1) in range of 10,0 to 300,0;
— overall dimensions, mm: 120 × 57 × 20;
— power supply — LiPo rechargeable battery, V: 3.7;
— range of operating temperatures for devices (basic): –20…+50 °C;
— continuous operating time, h: 8;
— device weight no more than, g: 150.
* Depends on characteristics of inspected material and type of transducer.
** Depends on characteristics of coating material.
Delivery set
Measuring unit with two transducers, one of them for inspection of metallic products under coating (additional transducers can be ordered), charging unit, operating manual, miniUSB — USB cable for connection with PC, CD with software for transfer to PC and processing of measuring results, shock-proof small-sized case.
Guarantee period
— measuring unit — 3 years;
— transducers — 2 years.
B6-C Leaflet to download below
B6-C
Probes for thickness gauge B6-C
Dual element probe 5A10
Purpose
Measurement of thickness of walls of metal tanks, containers, pipes of big diameters.
Basic technical characteristics
— operation frequency, MHz, 5;
— thickness range (for steel), mm, from 1.0 to 200.0;
— through paint testing (only for flat objects), (max thickness of paint ≤0,5 mm);
— dimensions of contact surface, mm, Ø10;
— overall dimensions, mm, Ø18 x 26;
— operating temperature range: from -10° to +60 °C.
Particulars
Utilization of new synthetic materials provides low wear and good interfacing with rough surfaces.
Dual element probe 10A6
Purpose
Measurement of thickness of pipes, curved areas of containers with small radius.
Basic technical characteristics
— operation frequency, MHz, 10;
— thickness range (for steel), mm, from 0.6 to 20;
— dimensions of contact surface, mm, Ø6;
— overall dimensions, mm, Ø14 x 26;
— operating temperature range: from -10° to +60 °C.
Particulars
Utilization of new synthetic materials provides low wear and good interfacing with rough surfaces.
Dual element probe 2.5A12
Purpose
Measurement of residual wall thickness of metallic and non-metallic products during manufacture, operation and repare.
Basic technical characteristics
— operation frequency, MHz, 2.5;
— thickness range (for steel), mm, from 3.0 to 200.0;
— dimensions of contact surface, mm, Ø15;
— overall dimensions, mm, Ø22 x 45;
— operating temperature range: from -10° to +60 °C.
Particulars
Good acoustic contact on rough and very rusted surface.
Small-sized dual element probe 10A4
Purpose
Measurement of wall thickness of pipes, thin metal plates.
Basic technical characteristics
— operation frequency, MHz, 10;
— thickness range (for steel), mm, from 0.5 to 10;
— dimensions of contact surface, mm, Ø4.5;
— overall dimensions, mm, Ø9 x 11;
— operating temperature range: from -10° to +60 °C.
Particulars
High locality, possibility of testing in hard-to-access locations (longerons, pipes with small diameter, etc).
Dual element high-temperature probe 5T12
Purpose
Intended for testing the thickness of hot metal parts. Used with thickness meter B6-С.
Basic technical characteristics
— operation frequency, MHz, 5;
— thickness range (for steel), mm, from 1.0 to 100.0;
— dimensions of contact surface, mm, Ø14;
— overall dimensions, mm, Ø26 x 48;
— operating temperature range: from -10° to +250 °C.
Through paint probes
Dual element probe 5E10
Purpose
Measuring of thickness of walls of metal tanks, conteiners, pipes of big diameters, also to read through and ignore coatings.
Basic technical characteristics
— operation frequency, MHz, 5;
— thickness range (for steel), mm, from 1.5 to 300.0 (in Pulse-Echo and Pulse-Echo-Coating); from 1.0 to 20.0 (in Echo-Echo mode);
— coating up to 2.0 mm (in Echo-Echo mode); coatings up to 1.0 mm (Pulse-Echo-Coating mode);
— dimensions of contact surface, mm, Ø12;
— overall dimensions, mm, Ø25 x 47;
— operating temperature range: from -10° to +60 °C.
Dual element probe 10E6
Purpose
Measurement of thickness of pipes, curved areas of containers with small radius, also to read through and ignore coatings.
Basic technical characteristics
— operation frequency, MHz, 10;
— thickness range (for steel), mm, from 0.6 to 50.0 (in Pulse-Echo-Mode); from 0.6 to 10.0 (in Echo-Echo mode, coatings up to 1.0 mm); from 0.6 to 25.0 (in Pulse-Echo-Coating mode, coatings up to 0.5 mm)
— dimensions of contact surface, mm, Ø8;
— overall dimensions, mm, Ø25 x 47;
— operating temperature range: from -10° to +60 °C.
Particulars
Good acoustic contact on rough and very rusted surface.
Delay line probe 15DL6
Purpose
It’s recommends for thickness measurement of very thin materials.
Basic technical characteristics
— operation frequency, MHz, 15;
— thickness range (for steel), mm, from 0.25 to 10.0;
— thirough paint range, mm, from 0.25 to 10.0 (up to 2 mm coating);
— dimensions of contact surface, mm, Ø7.4;
— overall dimensions, mm, Ø14 x 25.5.
Particulars
High locality, possibility of testing in hard-to-access locations (longerons, pipes with small diameter, etc).
Single element probes with protective membrane (for Echo-Echo mode)
P111-S5C
Purpose
Measurement of residual wall thickness of metallic and non-metallic products for use with the gauge in Echo-Echo mode or Multiple-Echo mode to read through and ignore coatings.
Basic technical characteristics
— operation frequency, MHz, 5;
— thickness range (for steel), mm, from 1.0 to 200.0;
— dimensions of contact surface, mm, Ø16;
— overall dimensions, mm, Ø26 x 46;
— operating temperature range: from -10° to +60 °C;
— protective membrane: polyurethane membrane.
Particulars
Good acoustic contact on rough and very rusted surfaces. Through coatings measuring.
P111-S2C
Purpose
Measurement of residual wall thickness of metallic and non-metallic products for use with the gauge in Echo-Echo mode to read through and ignore coatings.
Basic technical characteristics
— operation frequency, MHz, 2.25;
— thickness range (for steel), mm, from 2.0 to 150.0;
— dimensions of contact surface, mm, Ø16;
— overall dimensions, mm, Ø26 x 46;
— operating temperature range: from -10° to +60 °C;
— protective membrane: polyurethane membrane.
Particulars
Good acoustic contact on rough and very rusted surfaces. Through coatings measuring.