FxP, NFxP

FxP, NFxP

F and NF series probes for measuring thickness of internal pipe coatings.

Description

Probe for thickness measuring inside pipes made of ferromagnetic (FxP) and non-ferromagnetic (NFxP) materials.

Particulars
Design of probes allows control in manual mode the pipes opened from two sides. Appliances for pilling probe depends on diameter and length of pipe.

Features

  • measurement range, T, mm: 0-15;
  • minimum diameter of pipe, mm: 14;
  • pusher length, mm: 500-6000*.

* Pusher length is specified when ordering.

Basic technical characteristics

Probe Measurement range, T, mm Measurement accuracy, mm Minimum diameter of pipe, mm
F1xP20 0-0.3 ±(0.01T+0.001) 20
F2xP20 0-3 0-1 mm: ±(0.015T+0.001)
1-3 mm: ±0.02T
20
F2xP40 0-3 0-1 mm: ±(0.015T+0.001)
1-3 mm: ±0.02T
40
F3xP40 0-6 0-1 mm: ±(0.015T+0. 005)
1-6 mm: ±0.02T
40
NF1xP20 0-2 0-1 mm: ±(0.015T+0.001)
1-2 mm: ±0.02T
20
NF1xP40 0-2 0-1 mm: ±(0.015T+0.001)
1-2 mm: ±0.02T
40
NF2xP40 0-15 0-10 mm: ±(0.015T+0.01)
10-15 mm: ±0.02T
40